What are the guidelines for cleaning and sterilizing the Featured Shell?

Rinsing: Rinse contaminants from the Featured Shell with distilled water for at least 1 minute. 


Drying: Wipe the Featured Shell with a clean lint-free cloth and ensure its surface is dry. 


Sterilization: Seal the Featured Shell in a high-temperature and high-pressure sterilization bag and place it in a 

sterilizer for sterilization. (Sterilization parameters: Pre-vacuum type, temperature: 134℃, relative pressure: 

205.8 kPa, time: 4 minutes, drying time: 30 minutes; or lower exhaust type, temperature: 121℃, relative 

pressure: 102.9 kPa, time: 30 minutes, drying time: 30 minutes. The sterilization process must be vacuumed.)


Note: 


Please clean and sterilize the featured shell before initial use and after each use. 


The maximum number of sterilization cycles for the featured shell is 50.