How can we clean and sterilize the Jaw Motion Tracker?

Rinsing: After removing the Jaw Motion Tracker from the patient's mouth, peel off the markers, and immediately 

rinse it under running water for at least 1 minute to initially remove surface contaminants. Manual cleaning is 

recommended, and the recommended water temperature for cleaning is 15℃~30℃. 


Soaking: Immerse the rinsed Jaw Motion Tracker in an enzymatic cleaner. The recommended ratio of cleaner to 

water is 1:100-200, and soaking for at least 10 minutes. 


Cleaning: During the soaking process, gently stir the solution by "rotating" the liquid to further clean the Jaw 

Motion Tracker until the soaking is completed. 


Rinsing: Rinse the residual enzymatic cleaner on the Jaw Motion Tracker with distilled water for at least 1 minute 

to ensure there are no contaminants on the surface of the Jaw Motion Tracker. 


Drying: Wipe the Jaw Motion Tracker with a clean lint-free cloth and ensure its surface is dry. 


Sterilization: Seal the Jaw Motion Tracker in a high-temperature and high-pressure sterilization bag and place it in 

a sterilizer for sterilization. 

(Sterilization parameters: Pre-vacuum type, temperature: 134℃, relative pressure: 205.8 kPa, time: 4 minutes, drying time: 30 minutes; or lower exhaust type, temperature: 121℃, relative pressure: 102.9 kPa, time: 30 minutes, drying time: 30 minutes. The sterilization process must be vacuumed.)


Note:

Please clean and sterilize the Jaw Motion Tracker before initial use and after each use. 


The maximum number of sterilization cycles for the Jaw Motion Tracker is 100.