Rinsing: After removing the Jaw Motion Tracker from the patient's mouth, peel off the markers, and immediately
rinse it under running water for at least 1 minute to initially remove surface contaminants. Manual cleaning is
recommended, and the recommended water temperature for cleaning is 15℃~30℃.
Soaking: Immerse the rinsed Jaw Motion Tracker in an enzymatic cleaner. The recommended ratio of cleaner to
water is 1:100-200, and soaking for at least 10 minutes.
Cleaning: During the soaking process, gently stir the solution by "rotating" the liquid to further clean the Jaw
Motion Tracker until the soaking is completed.
Rinsing: Rinse the residual enzymatic cleaner on the Jaw Motion Tracker with distilled water for at least 1 minute
to ensure there are no contaminants on the surface of the Jaw Motion Tracker.
Drying: Wipe the Jaw Motion Tracker with a clean lint-free cloth and ensure its surface is dry.
Sterilization: Seal the Jaw Motion Tracker in a high-temperature and high-pressure sterilization bag and place it in
a sterilizer for sterilization.
(Sterilization parameters: Pre-vacuum type, temperature: 134℃, relative pressure: 205.8 kPa, time: 4 minutes, drying time: 30 minutes; or lower exhaust type, temperature: 121℃, relative pressure: 102.9 kPa, time: 30 minutes, drying time: 30 minutes. The sterilization process must be vacuumed.)
Note:
Please clean and sterilize the Jaw Motion Tracker before initial use and after each use.
The maximum number of sterilization cycles for the Jaw Motion Tracker is 100.