Accufab-CEL Printer Mini Platform Non-adhesion Issue

Leveling operation might be forgotten to take when users switch Accufab-CEL standard platform to mini platform, due to the different height between Accufab-CEL standard platform and mini platform, there there will be a gap below platform surface to resin tank film.

Accufab-CEL Standard Platform


Accufab-CEL Mini Platform



Please strictly to manual leveling every time when switch from Standard Platform to Mini Platform or reverse.


Platform Levelling for AccuFab-CEL : Dental Support

Did you find it helpful? Yes No