Situation:
Snaptool can connect to the controller (lower-level machine) normally, but operations experience noticeable delays, and the light uniformity distortion calibration consistently fails.
Cause:
Snaptool and the controller are connected via a wireless network with insufficient speed. During the light uniformity distortion calibration process, a large amount of data is exchanged between Snaptool and the controller, with extremely high requirements for real-time interaction to perform calculation and optimization. This process requires very high network speed, which is currently supported only by a direct wired network connection.
Main Solution:
Use a wired network to directly connect the upper-level and lower-level machines, then connect to the controller using Snaptool from the upper-level machine.