When should you level the platform of AccuFab-CEL?

Scene 1: First-Time Use of the Small Build Platform


Leveling Method: Full (Advanced) Leveling

Reason: The small platform has a different Z-height.

(Set to the same Z-axis position as the standard build platform)
Note: After this initial leveling, both the small and standard platforms can be used interchangeably.

 

Scene 2: Switching Between Standard and Small Platforms (After Initial Setup)


Leveling Method: No additional leveling required
Reason: If full leveling has already been done for both platforms, you can switch between them without re-leveling.


Scene 3: Print Failure Occurs Occasionally


Leveling Method: Quick Leveling
Reason: Minor shifts may occur over time or after frequent use. Quick leveling can correct small deviations.
Note: Only level the platform currently in use.

 

Scene 4: Frequent or Repeated Print Failures Even After Quick Leveling


Leveling Method: Full (Advanced) Leveling
Reason: Persistent failures may indicate more significant deviation.
Note:
 After full leveling, both platforms must be leveled again if you plan to use both.