Symptom:
Possible reason:
- The cleaning solution does the damage to the surface
- Ultrasonic cleaner does the damage to the surface.
Important Notice :
- Use water and enzymatic cleaning solution with proper ratio of 1:100-200.
- Recommended soak time of 10 mins and above.
- Wipe with lint-free cloth.
- Sterilization according to our official guidance.
Recommendation/Tested
Brand | Name | Product Code | Link | Reference |
3M | 3M Rapid Enzyme Cleaner | 70500 | 3M: https://www.3mindia.in/3M/en_IN/p/d/v000429544/ Amazon: https://www.amazon.in/3M-Rapid-Multi-Enzyme-Cleaner-70500/dp/B0BLNSK331 | ![]() |
3M | 3M™ 快速高效多酶清洗液 | 70503 | 3M: https://www.3m.com.cn/3M/zh_CN/p/d/v101084126/ | ![]() |
DO NOT USE
- Do not use cleaning solutions that contain alkali, oxidants, or any material sensitive to aluminum.
- Material, such as: Hydrogen peroxide and potassium hydroxide, are strictly forbidden.
- Do not use tap water to clean the coded scanbodies, instead, use distilled water.
- DO NOT use an ultrasonic cleaner for coded scanbodies. Which can damage the surfaces and leading to a loss of accuracy/dot capturing issue during scan.
Name | Reference |
Solution that are not for aluminum part | ![]() |
![]() | ![]() |
![]() | |
![]() | |
![]() | |
![]() |