Symptom:


Possible reason:
- The cleaning solution does the damage to the surface
- Ultrasonic cleaner does the damage to the surface.
Important Notice :
- Use distilled water and enzymatic cleaning solution with proper ratio of 1:100-200.
- Recommended soak time of 10 mins and above.
- Wipe with lint-free cloth.
- Sterilization according to our official guidance.
Recommendation/Tested
| Brand | Name | Product Code | Link | Reference |
| 3M | 3M Rapid Enzyme Cleaner | 70500 | 3M: https://www.3mindia.in/3M/en_IN/p/d/v000429544/ Amazon: https://www.amazon.in/3M-Rapid-Multi-Enzyme-Cleaner-70500/dp/B0BLNSK331 | ![]() |
| 3M | 3M™ 快速高效多酶清洗液 | 70503 | 3M: https://www.3m.com.cn/3M/zh_CN/p/d/v101084126/ | ![]() |
| Dr. Schumacher | PLURAZYME™ EXTRA | PLURAZYME EXTRA: PLURAZYME EXTRA | Dr. Schumacher | ![]() | |
| DR.WEIGERT | neodisher MediZym | https://www.drweigert.com/com/products-system-solutions/product/neodisher-medizym | ![]() |
DO NOT USE
- Do not use cleaning solutions that contain alkali, oxidants, or any material sensitive to aluminum.
- Material, such as: Hydrogen peroxide and potassium hydroxide, are strictly forbidden.
- Do not use tap water to clean the coded scanbodies, instead, use distilled water.
- DO NOT use an ultrasonic cleaner for coded scanbodies. Which can damage the surfaces and leading to a loss of accuracy/dot capturing issue during scan.
| Name | Reference |
| Solution that are not for aluminum part | ![]() |
![]() | ![]() |
![]() | |
![]() | |
![]() | |
![]() | |
| Rotagerm ultra | ![]() |
| Quitanet Ultra | ![]() |
| Virusolve+Concentrate | ![]()
![]() |
ANIOS Tri Enzymatic (This will make the markers fade.) | ![]() ![]() |
| DURR | ![]() |
| SACO | ![]() |
PS:
If you find that certain cleaning or disinfection solutions may damage the surface of the scanbody, or if you are willing to share information about the cleaning agents you are currently using, it would be of great help to us. Such feedback will not only assist us in improving our product usage recommendations, but can also be shared with a wider group of users to enhance overall safety and scanning performance.
Please provide the detailed information in the form below----Information Collection